Title: | Performance characterization of low-cost, high-speed, portable pulsed laser diode photoacoustic tomography (PLD-PAT) system | Authors: | Upputuri, Paul Kumar Pramanik, Manojit |
Issue Date: | 2015 | Source: | Upputuri, P. K., & Pramanik, M. (2015). Performance characterization of low-cost, high-speed, portable pulsed laser diode photoacoustic tomography (PLD-PAT) system. Biomedical Optics Express, 6(10), 4118-4129. | Series/Report no.: | Biomedical Optics Express | Abstract: | Photoacoustic tomography systems that uses Q-switched Nd:YAG/OPO pulsed lasers are expensive, bulky, and hence limits its use in clinical applications. The low pulse repetition rate of these lasers makes it unsuitable for real-time imaging when used with single-element ultrasound detector. In this work, we present a pulsed laser diode photoacoustic tomography (PLD-PAT) system that integrates a compact PLD inside a single-detector circular scanning geometry. We compared its performance against the traditional Nd:YAG/OPO based PAT system in terms of imaging depth, resolution, imaging time etc. The PLD provides near-infrared pulses at ~803 nm wavelength with pulse energy ~1.4 mJ/pulse at 7 kHz repetition rate. The PLD-PAT system is capable of providing 2D image in scan time as small as 3 sec with a signal-to-noise ratio ~30. High-speed and deep-tissue imaging is demonstrated on phantoms and biological samples. The PLD-PAT system is inexpensive, portable, allows high-speed PAT imaging, and its performance is as good as traditional expensive OPO based PAT system. Therefore, it holds promises for future translational biomedical imaging applications. | URI: | https://hdl.handle.net/10356/103440 http://hdl.handle.net/10220/38746 |
ISSN: | 2156-7085 | DOI: | 10.1364/BOE.6.004118 | Rights: | © 2015 Optical Society of America. This is the author created version of a work that has been peer reviewed and accepted for publication by Biomedical Optics Express, Optical Society of America. It incorporates referee’s comments but changes resulting from the publishing process, such as copyediting, structural formatting, may not be reflected in this document. The published version is available at: [http://dx.doi.org/10.1364/BOE.6.004118]. | Fulltext Permission: | open | Fulltext Availability: | With Fulltext |
Appears in Collections: | SCBE Journal Articles |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
BOE_paul_rev1_final_redLinedVersion.pdf | 447.54 kB | Adobe PDF | ![]() View/Open |
SCOPUSTM
Citations
5
104
Updated on Mar 26, 2023
Web of ScienceTM
Citations
5
99
Updated on Mar 27, 2023
Page view(s) 50
470
Updated on Mar 29, 2023
Download(s) 20
189
Updated on Mar 29, 2023
Google ScholarTM
Check
Altmetric
Items in DR-NTU are protected by copyright, with all rights reserved, unless otherwise indicated.