Please use this identifier to cite or link to this item: https://hdl.handle.net/10356/106904
Title: Hollow agarose microneedle with silver coating for intradermal surface-enhanced Raman measurements : a skin-mimicking phantom study
Authors: Yuen, Clement
Liu, Quan
Keywords: DRNTU::Science::Medicine::Biomedical engineering
Issue Date: 2015
Source: Yuen, C., & Liu, Q. (2015). Hollow agarose microneedle with silver coating for intradermal surface-enhanced Raman measurements : a skin-mimicking phantom study. Journal of biomedical optics, 20(6), 061102-.
Series/Report no.: Journal of biomedical optics
Abstract: Human intradermal components contain important clinical information beneficial to the field of immunology and disease diagnosis. Although microneedles have shown great potential to act as probes to break the human skin barrier for the minimally invasive measurement of intradermal components, metal microneedles that include stainless steel could cause the following problems: (1) sharp waste production, and (2) contamination due to reuse of microneedles especially in developing regions. In this study, we fabricate agarose microneedles coated with a layer of silver (Ag) and demonstrate their use as a probe for the realization of intradermal surface-enhanced Raman scattering measurements in a set of skin-mimicking phantoms. The Ag-coated agarose microneedle quantifies a range of glucose concentrations from 5 to 150 mM inside the skin phantoms with a root-mean-square error of 5.1 mM within 10 s. The needle is found enlarged by 53.9% after another 6 min inside the phantom. The shape-changing capability of this agarose microneedle ensures that the reuse of these microneedles is impossible, thus avoiding sharp waste production and preventing needle contamination, which shows the great potential for safe and effective needle-based measurements.
URI: https://hdl.handle.net/10356/106904
http://hdl.handle.net/10220/25242
ISSN: 1083-3668
DOI: 10.1117/1.JBO.20.6.061102
Schools: School of Chemical and Biomedical Engineering 
Rights: © 2015 Society of Photo-Optical Instrumentation Engineers. This paper was published in Journal of Biomedical Optics and is made available as an electronic reprint (preprint) with permission of Society of Photo-Optical Instrumentation Engineers. The paper can be found at the following official DOI: [http://dx.doi.org/10.1117/1.JBO.20.6.061102].  One print or electronic copy may be made for personal use only. Systematic or multiple reproduction, distribution to multiple locations via electronic or other means, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper is prohibited and is subject to penalties under law.
Fulltext Permission: open
Fulltext Availability: With Fulltext
Appears in Collections:SCBE Journal Articles

Files in This Item:
File Description SizeFormat 
JBO_20_6_061102.pdf3.73 MBAdobe PDFThumbnail
View/Open

SCOPUSTM   
Citations 20

25
Updated on Mar 6, 2025

Web of ScienceTM
Citations 20

19
Updated on Oct 25, 2023

Page view(s) 20

758
Updated on Mar 20, 2025

Download(s) 20

337
Updated on Mar 20, 2025

Google ScholarTM

Check

Altmetric


Plumx

Items in DR-NTU are protected by copyright, with all rights reserved, unless otherwise indicated.