Title: A method to produce high contrast vein visualization in active dynamic thermography (ADT)
Authors: Saxena, Ashish
Ng, Eddie Yin Kwee
Canchi, Tejas
Lim, Jia Ler
Beruvar, Ayush Singh
Keywords: Engineering::Mechanical engineering
Issue Date: 2021
Source: Saxena, A., Ng, E. Y. K., Canchi, T., Lim, J. L. & Beruvar, A. S. (2021). A method to produce high contrast vein visualization in active dynamic thermography (ADT). Computers in Biology and Medicine, 132, 104309-. https://dx.doi.org/10.1016/j.compbiomed.2021.104309
Journal: Computers in Biology and Medicine
Abstract: In this study, a method that will aid in the visualization of vein topology on a target area on the body of a human subject is demonstrated. An external cooling means is configured to cool the left forearm of fourteen study participants, effecting an active thermal change or recovery in the target area upon removal of cooling. An infrared (IR) thermal camera was used to capture a series of transient thermal images. These images were then processed to extract Dynamic synthetic images (SI) throughout the active thermal change or recovery process. Dynamic SI was calculated using a quantitative parameter called tissue activity ratio (TAR), which is defined by the rate of rewarming to the rate of cooling at each pixel of interest. A fixed step size of rewarming temperature (0.5 °C) was used to progressively extract multiple synthetic images throughout the whole recovery process. Compared to a Static SI extraction method, where only a single SI results from the whole active dynamic thermography (ADT) sequence, this study demonstrates a live feed of high contrast vein visualizations by using the Dynamic SI method. Furthermore, the dependency of Dynamic SI contrast on the temperature of the external cooling stimulation was investigated. Three cooling stimulation temperatures (5 °C, 8 °C, and 11 °C) were tested, where no statistically significant difference in the resulting SI contrast was found. Lastly, a discussion is put forth on assisting venipuncture or cannulation-based clinical applications, through the incorporation of the proposed method with a projection system.
URI: https://hdl.handle.net/10356/156996
ISSN: 0010-4825
DOI: 10.1016/j.compbiomed.2021.104309
Schools: School of Mechanical and Aerospace Engineering 
Rights: © 2021 Elsevier Ltd. All rights reserved. This paper was published in Computers in Biology and Medicine and is made available with permission of Elsevier Ltd.
Fulltext Permission: open
Fulltext Availability: With Fulltext
Appears in Collections:MAE Journal Articles

SCOPUSTM   
Citations 50

6
Updated on Sep 23, 2023

Web of ScienceTM
Citations 50

6
Updated on Sep 22, 2023

Page view(s)

66
Updated on Sep 23, 2023

Download(s) 50

45
Updated on Sep 23, 2023

Google ScholarTM

Check

Altmetric


Plumx

Items in DR-NTU are protected by copyright, with all rights reserved, unless otherwise indicated.