Title: Experimental validation of tangential resolution improvement in photoacoustic tomography using modified delay-and-sum reconstruction algorithm
Authors: Kalva, Sandeep Kumar
Pramanik, Manojit
Keywords: Image reconstruction
Tangential resolution
Issue Date: 2016
Source: Kalva, S. K., & Pramanik, M. (2016). Experimental validation of tangential resolution improvement in photoacoustic tomography using modified delay-and-sum reconstruction algorithm. Journal of Biomedical Optics, 21(8), 086011-.
Series/Report no.: Journal of Biomedical Optics
Abstract: For a circular scanning geometry in photoacoustic tomography, the axial/radial resolution is spatially invariant and is not affected by the ultrasound transducer (UST, detector) aperture. However, the tangential resolution is dependent on the detector aperture size and it varies spatially. Many techniques were proposed to improve the tangential resolution, such as attaching a concave lens in front of the nonfocused transducer or using a virtual point detector. Both of these methods have difficulties. Therefore, a modified delay-and-sum reconstruction algorithm has been proposed which can be used together with a standard ultrasound detector (nonfocused) to improve the tangential resolution. In this work, we validate the modified delay-and-sum algorithm experimentally for both flat and cylindrically focused USTs. More than threefold improvement in tangential resolution is observed. It is also shown that the object shape is recovered with this modified algorithm, which is very helpful for diagnosis and treatment purposes.
URI: https://hdl.handle.net/10356/82178
http://hdl.handle.net/10220/41172
ISSN: 1083-3668
DOI: 10.1117/1.JBO.21.8.086011
Schools: School of Chemical and Biomedical Engineering 
Rights: © 2016 Society of Photo-optical Instrumentation Engineers (SPIE). This paper was published in Journal of Biomedical Optics and is made available as an electronic reprint (preprint) with permission of SPIE. The published version is available at: [http://dx.doi.org/10.1117/1.JBO.21.8.086011]. One print or electronic copy may be made for personal use only. Systematic or multiple reproduction, distribution to multiple locations via electronic or other means, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper is prohibited and is subject to penalties under law.
Fulltext Permission: open
Fulltext Availability: With Fulltext
Appears in Collections:SCBE Journal Articles

SCOPUSTM   
Citations 5

64
Updated on Jun 17, 2024

Web of ScienceTM
Citations 5

51
Updated on Oct 25, 2023

Page view(s) 50

479
Updated on Jun 24, 2024

Download(s) 20

244
Updated on Jun 24, 2024

Google ScholarTM

Check

Altmetric


Plumx

Items in DR-NTU are protected by copyright, with all rights reserved, unless otherwise indicated.