Please use this identifier to cite or link to this item: https://hdl.handle.net/10356/87624
Title: Micro-optical coherence tomography (μOCT) in vivo
Authors: Cui, Dongyao
Liu, Xinyu
Yu, Xiaojun
Shum, Ping
Liu, Linbo
Keywords: Optical Coherence Tomography
General
DRNTU::Engineering::Electrical and electronic engineering
Issue Date: 2014
Source: Cui, D., Liu, X., Yu, X., Shum, P., & Liu, L. (2014). Micro-optical coherence tomography (μOCT) in vivo. Imaging Systems and Applications 2014, IM3C.5-. doi:10.1364/ISA.2014.IM3C.5
Conference: Imaging Systems and Applications 2014
Abstract: We have developed a Micro- optical coherence tomography (μOCT) for in vivo use. We extended line width of the light source and have achieved a longitudinal resolution of 1.9μm in free space and 1.38μm in tissue. In vivo imaging was conducted using zebra fish larvae and cellular structures such as red blood cells in tail artery can be observed.
URI: https://hdl.handle.net/10356/87624
http://hdl.handle.net/10220/46796
DOI: 10.1364/ISA.2014.IM3C.5
Schools: School of Chemical and Biomedical Engineering 
School of Electrical and Electronic Engineering 
Research Centres: Photonics Centre of Excellence 
Rights: © 2014 The Author(s) Optical Society of America(OSA). This paper was published in Imaging Systems and Applications 2014 and is made available as an electronic reprint (preprint) with permission of The Author(s) Optical Society of America(OSA). The published version is available at: [http://dx.doi.org/10.1364/ISA.2014.IM3C.5]. One print or electronic copy may be made for personal use only. Systematic or multiple reproduction, distribution to multiple locations via electronic or other means, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper is prohibited and is subject to penalties under law.
Fulltext Permission: open
Fulltext Availability: With Fulltext
Appears in Collections:EEE Conference Papers

Files in This Item:
File Description SizeFormat 
Micro-optical coherence tomography (μOCT) in vivo.pdf696.55 kBAdobe PDFThumbnail
View/Open

Page view(s)

477
Updated on Mar 27, 2025

Download(s) 20

245
Updated on Mar 27, 2025

Google ScholarTM

Check

Altmetric


Plumx

Items in DR-NTU are protected by copyright, with all rights reserved, unless otherwise indicated.