Title: Spatially variant regularization based on model resolution and fidelity embedding characteristics improves photoacoustic tomography
Authors: Kalva, Sandeep Kumar
Pramanik, Manojit
Yalavarthy, Phaneendra K.
Sanny, Dween Rabius
Prakash, Jaya
Keywords: Photoacoustic Tomography
Image Reconstruction
DRNTU::Engineering::Chemical engineering
Issue Date: 2018
Source: Sanny, D. R., Prakash, J., Kalva, S. K., Pramanik, M. & Yalavarthy, P. K. (2018). Spatially variant regularization based on model resolution and fidelity embedding characteristics improves photoacoustic tomography. Journal of Biomedical Optics, 23(10), 1-. doi:10.1117/1.JBO.23.10.100502
Series/Report no.: Journal of Biomedical Optics
Abstract: Photoacoustic tomography tends to be an ill-conditioned problem with noisy limited data, requiring imposition of regularization constraints like standard Tikhonov or total-variation to reconstruct meaningful initial pressure rise distribution from the tomographic acoustic measurements acquired at the boundary of the tissue. However, these regularization schemes does not account for non-uniform sensitivity arising due to limited detector placement at the boundary of tissue as well as other system parameters. For the first time, in this work, two regularization schemes were developed within the Tikhonov framework to address these issues in photoacoustic imaging. The model-resolution, based spatially varying regularization, and fidelity embedded regularization, based on orthogonality between the columns of system matrix were introduced in this work. These were systematically evaluated with the help of numerical and in-vivo mice data. It was shown that the performance of the proposed spatially varying regularization schemes were superior (with atleast 2 dB or 1.58 times improvement in the SNR) compared to standard Tikhonov/total-variation based regularization schemes.
URI: https://hdl.handle.net/10356/90005
http://hdl.handle.net/10220/46722
ISSN: 1083-3668
DOI: 10.1117/1.JBO.23.10.100502
Schools: School of Chemical and Biomedical Engineering 
Rights: © 2018 Society of Photo-optical Instrumentation Engineers (SPIE). This is the author created version of a work that has been peer reviewed and accepted for publication by Journal of Biomedical Optics, Society of Photo-optical Instrumentation Engineers (SPIE). It incorporates referee’s comments but changes resulting from the publishing process, such as copyediting, structural formatting, may not be reflected in this document. The published version is available at: [http://dx.doi.org/10.1117/1.JBO.23.10.100502].
Fulltext Permission: open
Fulltext Availability: With Fulltext
Appears in Collections:SCBE Journal Articles

Files in This Item:
File Description SizeFormat 
article.pdf1.02 MBAdobe PDFThumbnail
View/Open

SCOPUSTM   
Citations 50

5
Updated on May 7, 2025

Web of ScienceTM
Citations 20

4
Updated on Oct 30, 2023

Page view(s) 50

584
Updated on May 6, 2025

Download(s) 50

123
Updated on May 6, 2025

Google ScholarTM

Check

Altmetric


Plumx

Items in DR-NTU are protected by copyright, with all rights reserved, unless otherwise indicated.