Please use this identifier to cite or link to this item: https://hdl.handle.net/10356/94592
Title: Passive micromixer for luminol-peroxide chemiluminescence detection
Authors: Lok, Khoi Seng
Kwok, Yien Chian
Nguyen, Nam-Trung
Keywords: DRNTU::Engineering::Mechanical engineering
Issue Date: 2011
Source: Lok, K. S., Kwok, Y. C., & Nguyen, N. T. (2011). Passive Micromixer for Luminol-peroxide Chemiluminescence Detection. Analyst, 136, 2586-2591.
Series/Report no.: Analyst
Abstract: This paper reports a microchip with an integrated passive micromixer based on chaotic advection. The micromixer with staggered herringbone structures was used for luminol-peroxide chemiluminescence detection. The micromixer was examined to assess its suitability for chemiluminescence reaction. The relationship between the flow rate and the location of maximum chemiluminescence intensity was investigated. The light intensity was detected using an optical fiber attached along the mixing channel and a photon detector. A linear correlation between chemiluminescence intensity and the concentration of cobalt(II) ions or hydrogen peroxide was observed. This microchip has a potential application in environmental monitoring for industries involved in heavy metals and in medical diagnostics.
URI: https://hdl.handle.net/10356/94592
http://hdl.handle.net/10220/7770
DOI: 10.1039/c1an15280g
Schools: School of Mechanical and Aerospace Engineering 
Rights: © 2011 The Royal Society of Chemistry. This is the author created version of a work that has been peer reviewed and accepted for publication by Analyst, The Royal Society of Chemistry 2011. It incorporates referee’s comments but changes resulting from the publishing process, such as copyediting, structural formatting, may not be reflected in this document. The published version is available at: [DOI: http://dx.doi.org/10.1039/c1an15280g].
Fulltext Permission: open
Fulltext Availability: With Fulltext
Appears in Collections:MAE Journal Articles

Files in This Item:
File Description SizeFormat 
Passive Micromixer for Luminol.pdf184.87 kBAdobe PDFThumbnail
View/Open

SCOPUSTM   
Citations 20

21
Updated on Mar 22, 2025

Web of ScienceTM
Citations 20

20
Updated on Oct 29, 2023

Page view(s) 20

786
Updated on Mar 24, 2025

Download(s) 5

575
Updated on Mar 24, 2025

Google ScholarTM

Check

Altmetric


Plumx

Items in DR-NTU are protected by copyright, with all rights reserved, unless otherwise indicated.