Please use this identifier to cite or link to this item: https://hdl.handle.net/10356/98121
Title: Microfluidic rheometer based on hydrodynamic focusing
Authors: Nguyen, Nam-Trung
Yap, Yit Fatt
Sumargo, Agung
Keywords: DRNTU::Engineering::Mechanical engineering
Issue Date: 2008
Source: Nguyen, N. T., Yap, Y. F., & Sumargo, A. (2008). Microfluidic Rheometer Based on Hydrodynamic Focusing. Measurement Science and Technology, 19(8).
Series/Report no.: Measurement science and technology
Abstract: This paper reports the concept and the optimization of a microfluidic rheometer based on hydrodynamic focusing. In our microfluidic rheometer, a sample stream is sandwiched between two sheath streams. The width of the middle stream depends on the viscosity ratio and the flow rate ratio of the liquids involved. Fixing the flow rate ratio and using a known Newtonian liquid for the sheath streams, the viscosity and the shear stress of the sample stream can be determined by measuring its width and using a prediction algorithm that uses the known channel geometries, fluid properties, flow rates and the focused width as input parameters. The optimization reveals that a measurement channel with a high aspect ratio is more suitable for a sample liquid with viscosity higher than the reference value. For a sample liquid with viscosity of the same order of magnitude or lower than the reference, a low aspect ratio is more suitable. Furthermore, the measurement range and the relative error can be improved by adjusting the flow rate ratio between the core stream and the sheath stream. A microfluidic device was fabricated in polymethylmethacrylate (PMMA). Using this device, viscosities of deionized (DI) water and polyethylene oxide (PEO) solutions were measured and compared with results obtained from a commercial rheometer.
URI: https://hdl.handle.net/10356/98121
http://hdl.handle.net/10220/7856
DOI: 10.1088/0957-0233/19/8/085405
Schools: School of Mechanical and Aerospace Engineering 
Rights: © 2008 IOP Publishing Ltd. This is the author created version of a work that has been peer reviewed and accepted for publication by Measurement Science and Technology , IOP Publishing Ltd. It incorporates referee’s comments but changes resulting from the publishing process, such as copyediting, structural formatting, may not be reflected in this document. The published version is available at: [http://dx.doi.org.ezlibproxy1.ntu.edu.sg/10.1088/0957-0233/19/8/085405].
Fulltext Permission: open
Fulltext Availability: With Fulltext
Appears in Collections:MAE Journal Articles

Files in This Item:
File Description SizeFormat 
ViscometerPaperHighlighted.pdf641.46 kBAdobe PDFThumbnail
View/Open

SCOPUSTM   
Citations 10

34
Updated on Mar 5, 2025

Web of ScienceTM
Citations 10

33
Updated on Oct 26, 2023

Page view(s) 5

1,116
Updated on Mar 15, 2025

Download(s) 10

472
Updated on Mar 15, 2025

Google ScholarTM

Check

Altmetric


Plumx

Items in DR-NTU are protected by copyright, with all rights reserved, unless otherwise indicated.